OxyScan
Graphic
UMS Oxygen Sensor 501®
Users Manual
Introduction
Thank you for choosing the UMS Micro Oxygen Sensor 501®!
With the UMS Micro Oxygen Sensor 501® you have purchased a highly innovative
product.
Please read these instructions carefully to avoid any possible damage to the sensor and
measuring instrument.
1. Components supplied
- measuring unit
- micro oxygen sensor
- calibration chamber
- power supply for charging the integrated accumulator
- transport case
- this manual
Carefully unpack the components and check that everything is present.
2. Handling instructions
The heart of the oxygen sensor is the highly sensitive sensor tip.
When the sensor tip of the oxygen electrode is accidentally knocked, the membrane at the
tip can be damaged. Constantly increasing display figures and overflow indicate a damaged
membrane. In this case the sensor must be replaced.
If there are any problems, please contact us.
3. Explanation of operation

1 On / Off - switch
2 Calibration key
3 Menu/Enter: for confirmation of choice or entering menu
4 "Up": key for choosing up in the menu
5 "Down": key for choosing down in the menu
6 F - Key: only for customer-specific options, if need look at additions sheet
4. Operation
On top of the measuring unit are the connectors
- for the sensor ( connector in the middle - 5 pins ),
- for the power supply/charger ( left connector 7 pins )

Connect the plug of the sensor with the connector of the measuring unit (connector in
the middle - 5 pins).
For charging the integrated accumulators connect the power supply with the charging
connector top left of the measuring unit. The charging circuit stops charging
automatically if the accumulator has reached its full capacity. We recommend
switching off the unit while charging. The power supply can stay connected to keep the
accumulator fully charged.
If the red LED on the keyboard is on, the accumulator will be charged. If it is ready the
LED will blink.
5. Calibration
For calibrating put the sensor into the calibration chamber and press the "CAL."-key on the measuring unit. The calibration runs automatically and stops when the calibration value is stable. On the bottom end of the calibration chamber there is a black plug containing a sponge. This sponge must be kept damp approximately every 4 weeks as follows:
remove the black plug
dampen sponge with distilled water
close the calibration chamber
after about 15 minutes calibration can begin
The sensor should be stored into the calibration chamber. The sponge must be damp at all tries.
Hint:
Your oxygen measurement system is an highly innovative product. Your contribution for
exact measuring is the calibration of the system.
Frequent calibration increases accuracy of measurement!
6. Measurement
After calibrating the measurement system, measurement of the media can begin.
Hint:
You should place the sensor tip in middle depth of the vessel ( minimum 5 cm below the
surface).
The sensor need a minimum flow or a constant flow. Please consider this if installing the
sensor in a system.
7. Description of the display in measuring mode

1: oxygen content in % saturation
2: oxygen content in mg/l
3: temperature in °C
4: air pressure in hPa
5: diagram shows state of the integrated accumulator ( 0-100 % )
6: diagram of the last 30 values measured in % saturation ( 0-160 %sat )
7: diagram of the last 30 values measured in mg/l ( 0-16 mg/l )
8: diagram of the last 30 temperature values measured ( 0-48 °C )
9: diagram shows state of measurement memory ( 0-5000 values )
10: actual time
You can enter the menu by pressing "MENU/ENTER"-key. Then you can select
submenus described in the next sections.
To select a submenu use "Up"- or "Down"- keys and then press the
"MENU/ENTER"-key to enter it.
If you select "BACK" in the menu you will go back to the measuring mode. The
selection in the submenus works similarly like this.
8.1. Settings
In this menu you can make the following selections:
Set time: Set time as follows:
1. Move cursor with the arrow keys to the number to set
2. press MENU/ENTER"
3. Change the number with arrow keys
4. press MENU/ENTER"
5. go back to point 1 or select "BACK" and press "MENU/ENTER"
NTC-correction: Set temperature correction value with the arrow keys and press "MENU/ENTER". This value should not to be changed by the user.
Press.-correct.: Set air pressure correction value with the arrow keys and press "MENU/ENTER". This value is should not to be changed by the user.
Cal. End value: Set calibration end value with arrow keys and press "MENU/ENTER". Change this value only when an other calibration method is used. ( Standard value is 102.3 %sat )
Auto power off: Set the auto-power-off time with the arrow keys and press "MENU/ENTER". If there is no key pressed for the time adjusted here the device will switch off automatically. If you set the time to 0 the device will not switch off automatically.
Set language: You can switch between German and English display text. This manual refers to English texts.
Device information: Shows information about the device: serial numbers of device and sensor, Firmware version and some information for service and diagnostics use.
Hints:
The values for NTC-correction depends on the sensor and is set in our company. This value
is normally only to be changed if the sensor is replaced.
The value for pressure correction should never be changed under normal circumstances.
8.2. Log functions
Select this submenu for data logging.
In this menu you can make the following selections:
Set interval: Set the time between logging single values.
Start log now: Starts logging immediately.
Set start time: Set the time the logging starts. Setting this time should be done like "set time" described in section 7.1.
Stop logging: Logging will be stopped.
The last 3 entries will only be shown if convenient. (e.g. "Stop logging" will only be shown if there is any logging running).
Hint:
Data logging will also be continued if the device is switched off. Please note the
higher power consumption and shorter running time. The shorter the interval is set, the
faster the accumulator will discharge. At all selectable intervals a full accumulator will
be sufficient for filling the measurement memory completely.
8.3. Values
Select this submenu for watching or transmitting the logged values.
In this menu you can make the following selections:
Show values: The logged values will be shown as numbers. Use the arrow keys to select a value. Press "MENU/ENTER" for changing this mode.
Show Diagram: The logged values will be shown in a diagram. Select the value that the user requires.
Send values: The logged values will be sent by the infrared transceiver. See section 8.3.1 for details.
Delete values: All values will be deleted. ( only shown if there are values logged )
8.3.1. Data transmission
The OxyScan graphic can send data via its infrared interface to a PC or printer which must meet the following specifications:
19,2 kBps, 8 Databits, no Parity, 1 Stop bit ( = "8N1" ), 3/16 IR modulation
The optional infrared-adapter enables the sending of measured values to devices not
meeting this specifications which have a standard RS232 interface like most PCs and
some printers.
The data are sent as table in text format, so therefor can be easily sent directly to a
printer.
For data transmissions to a Windows-PC you can use the terminal software included with
most windows-versions. The values are separated by tabulator and the data rows are
separated by "NewLine", so they can be imported into most spreadsheet
applications (like MS Excel).
Please ensure that no objects are between the OxyScan and the Infrared receiver. If the
distance between the devices is longer then 50 cm they must be lined up. If they are
exactly lined up distances up to 5 m can be reached.
Attention:
Never look directly into the dark red infrared windows, direct eye
contact can cause retinal damage.
The oxygen sensor will be delivered in the calibration chamber. It is very important
that the calibration chamber is always kept damp. (look at point 5
"Calibration")
If the calibration chamber is not regularly used then store the sensor in an other vessel
filled with water. The sensor tip must be always in water.
10. Additional functions / Firmware update
Depending on the options you have chosen the microprocessor device can be equipped with
additional functions which are not used in this manual. These functions will be described
on a extra sheet.
Your OxyScan graphic can be equipped with further functions. Because microcontroller
technology is used we can realise many additional functions. Simply ask us!
All functions of the device are controlled by a microcontroller. On this microcontroller a
program developed by us is run. This program ( = firmware ) is saved in a memory module
that should only be replaced by us or other authorised persons.
Vigorous tests are carried out on the firmware. Nevertheless updating of the firmware is a
continuos process. When an update is needed then call us and tell us your current version
number ( look at 8.1. "device information" ). We will then inform you of the new
additional functions of the new firmware.
The oxygen sensor functions according the Clark Principle. It measures the oxygen
dissolved in the water. The oxygen diffuses through a membrane at the sensor tip and is
reduced at the cathode. The electrons generated at the anode lead to an very low electric
current, which is electronically monitored by the unit.
However, the current in the sensor is not only determined by the waters oxygen content,
but also by the temperature of the water and the air pressure. To compensate these factors
the measuring unit of the oxygen sensor is equipped with automatic temperature and air
pressure compensations.
Technical changes are reserved.
We undertake no liability whatsoever for damages arising from misuse and erroneous
application.
The entire measurement system carries a legal warranty of 6 months. The measuring unit
itself is covered by a further 6 months warranty.
On claims under the warranty please return both the sensor and measuring unit.
13. Technical specifications
measuring ranges
concentration range: 0.0 19.99 mg/l
resolution: 0.01 mg/l
accuracy: ± 0.1 mg/l
saturation range: 0.0 199.9 %sat
resolution: 0.1 %sat
accuracy: ± 0.5%
temperature range: 0.0 50.0 °C
resolution: 0.1 K
accuracy: 0.5 K
air pressure range: 800 1150 mbar
resolution: 1 mbar
accuracy: 10 mbar
temperature compensation automatically, range: 0 - 50 °C
air pressure compensation automatically, range: 800 - 1150 mbar
environmental temperature 0°C - 70°C
(measurement device)
display graphic-capable LCD 128 x 64 Points
electrode Pt-cathode, Ag-anode
power supply integrated accumulator 3,6 V / 2.1 Ah, NiMH
dimensions device 223 x 105 x 40 ( lxwxh, mm )
dimensions sensor 200 x 4.7 ( lxÆ , mm )
weight device approx. 500 g
weight sensor approx. 20 g
polarisation voltage 750 mV
polarisation time 5 min maximum ( 30 sec. 1 % )
sensor reaction time approx. 30 s
The lifespan of the electrode is approximately 2 years. Because the sensor is miniaturised it can not be regenerated by the user. This eliminates a number of error sources which result from the regeneration process.
14. Appendix
(The following tables are only valid under the mentioned conditions!)
Oxygen content in deionised water
in relation to temperature under normal pressure ( 1031 hPa = 760 Torr )

Maximum dissolved oxygen in freshwater
in mg/l at various temperatures and under normal pressure ( 1013 hPa = 760 Torr )
°C |
mg/l |
°C |
mg/l |
°C |
mg/l |
°C |
mg/l |
0 |
14.64 |
10 |
11.25 |
20 |
9.08 |
30 |
7.55 |
1 |
14.23 |
11 |
10.99 |
21 |
8.90 |
31 |
7.42 |
2 |
13.83 |
12 |
10.75 |
22 |
8.73 |
32 |
7.30 |
3 |
13.45 |
13 |
10.51 |
23 |
8.57 |
33 |
7.18 |
4 |
13.09 |
14 |
10.28 |
24 |
8.41 |
34 |
7.06 |
5 |
12.75 |
15 |
10.06 |
25 |
8.25 |
35 |
6.94 |
6 |
12.42 |
16 |
9.85 |
26 |
8.11 |
36 |
6.83 |
7 |
12.11 |
17 |
9.64 |
27 |
7.96 |
37 |
6.72 |
8 |
11.81 |
18 |
9.45 |
28 |
7.82 |
38 |
6.61 |
9 |
11.53 |
19 |
9.26 |
29 |
7.69 |
39 |
6.51 |
10 |
11.25 |
20 |
9.08 |
30 |
7.55 |
40 |
6.41 |
Oxygen content in seawater
in relation to salt content ( density ) and temperature
under normal pressure ( 1013 hPa = 760 Torr )
density in g/cm3 |
10°C O2 in mg/l |
15°C O2 in mg/l |
20°C O2 in mg/l |
25°C O2 in mg/l |
30°C O2 in mg/l |
1.000 |
11.25 |
10.06 |
9.09 |
8.26 |
7.49 |
1.005 |
10.81 |
9.69 |
8.76 |
7.96 |
7.21 |
1.010 |
10.38 |
9.32 |
8.44 |
7.65 |
6.93 |
1.015 |
9.94 |
8.95 |
8.11 |
7.35 |
6.65 |
1.020 |
9.51 |
8.58 |
7.79 |
7.05 |
6.38 |
1.025 |
9.07 |
8.21 |
7.46 |
6.74 |
6.10 |
1.030 |
8.64 |
7.85 |
7.14 |
6.44 |
5.82 |
Salt content ( salinity in g/l = ) of seawater
in relation to temperature an density of seawater
density in g/cm3 |
10°C salinity |
15°C salinity |
20°C salinity |
25°C salinity |
30°C salinity |
1.000 |
0 |
0 |
0 |
0 |
0 |
1.005 |
6.7 |
7.5 |
8.5 |
10.8 |
13.1 |
1.010 |
13.2 |
14.4 |
15.6 |
17.2 |
18.8 |
1.015 |
19.6 |
20.9 |
22.1 |
23.6 |
25.1 |
1.020 |
26.0 |
27.3 |
28.6 |
30.6 |
32.6 |
1.025 |
32.5 |
33.9 |
35.3 |
37.1 |
38.9 |
1.030 |
38.8 |
40.4 |
42.0 |
44.2 |
46.4 |
For technical queries and orders of accessories please contact:
AquaTech
Unterbrunnweg 3, A-6370 Kitzbühel/Austria
Tel: +43/5356-71399, Fax: +43/5356-64870
http://www.aquatech.8m.com